Specifications
Specifications describe the instrument's warranted performance. Non-warranted values are described as typical. All specifica-
tions apply after a 30-minute warm-up phase with 50 W source/load resistance. All specifications are valid from 0°C to 55°C
ambient temperature.
8114A Specifications
Timing Characteristics
Measured at 50% amplitude at fastest transitions in continuous mode and 50
WW
source impedance.
Width: Can be set as width, duty
cycle, or trailing edge delay.
Burst Count: 2 to 65536 (single or
double pulses).
Double Pulse: Two pulses are
generated per pulse period. The
first pulse starts at the start of
the period, the double-delay sets
the delay to the start of the
second pulse. Double pulses are
available in all trigger modes
except External Width.
Delay: Delay, phase or % of period.
Double pulse delay: Double pulse
and delay are mutually exclusive.
The delay between double pulses
can be set as delay or % of period
(minimum period 134 ns)
Duty cycle: Set between 0.1% and
100% (subject to width limits and
period specifications).
Figure 1 shows the specified
operating range for actual duty
cycle and amplitude. Note that in
double pulse mode the actual
duty cycle is twice the displayed
value.
Repeatability: Is typically four
times better than accuracy.
Commenti su questo manuale