Level/Pulse Performance Characteristics
Agilent 81101A
Amplitude
1.00 V to 50.0V or 20.0
mA to 1
.00A
(values approximately double for
source or load resistance > 1 k
W
).
Accuracy
±1 % of amplitude ± 100 mV
Resolution
3 digits, best case 10 mV
Baseline
Standard Instrument
Option 001
< 100
mV + 0.5% of amplitude.
25.0
V to 25.0V
[1]
Accuracy
±1% ±100
mV ±0.5% of amplitude.
Pulse Window
±50
V
Base line noise
10 mV RMS typ.
Output connectors
BNC
Source Impedance
Selectable 50
W
or High Impedance
(>1 k
W)
Short circuit current
2 A
Overshoot/preshoot/ringing
<± 5% of amplitude ±
100
mV
Settling time
<100
ns typical
Agilent 8114A
Level parameters: Can be entered
as voltage or current, as high and
low level, or as offset and ampli-
tude.
Load compensation: The actual
load value can be entered (for
loads differnt to 50 W) to display
actual output values.
On/off: Relays connect/disconnect
output.
Polarity: Positive or negative
pulses selectable.
Output Protection: Maximum extern-
al voltage 100 Vpp from 50 W
source (±10 Vdc from 0 W source).
Limit: Programmable level and
duty cycle limits restrict the
available output ranges to protect
the device-under-test.
Note: [1] 50 W source must be selected for variable baseline
Variable Baseline
-50
-25
-25
-5
+1 +25
+50
-1
Amplitude
V
V (Option 001)
Maximum Duty Cycle 15%
Maximum Duty Cycle, see Figure 2
0 V
0 V
+25
0 V
0 V
+5
Figure 2
Vedere la pagina 3