Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 31
strip, and the center conductors of each SMA
connector.
b Ensure that there is continuity between each of the
two pins on the right side of the pin strip, and the
SMA connector frames.
c Ensure that there is NO continuity between the SMA
connector center conductor and the SMA connector
frame (ground).