Agilent Technologies 8133A Manuale di Servizio Pagina 47

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 46
Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 47
3 Set the sampling mode.
a Select the Sampling tab of the Analyzer Setup dialog.
b Select State - Synchronous Sampling.
c Set the Trigger Position to 100% Poststore.
d Set the Acquisition Depth to 128K.
e Clear the Timing Zoom check box to turn Timing Zoom
off.
f Ensure that the sampling speed is set to 250 MHz in
the Sampling Options box.
Vedere la pagina 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 187 188

Commenti su questo manuale

Nessun commento