Agilent Technologies B1500A Manuale di Servizio Pagina 85

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 110
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 84
Module 6
Low Current Measurement
6-13
Change SMU Range to
AUTO or LIMITED 10 pA
Change Integration Time to
16 PLC
ZERO CHECK
Range and Integration Time
You must change the range setting from the default of LIMITED 1 nA to LIMITED 10 pA or to
AUTO. Otherwise you will have 100 fA resolution instead of the best 1 fA resolution possible.
Also change the High Resolution ADC Mode to PLC and Factor to 16. You will see +/- 3 fA
variation measurement result as shown in the previous slide.
Vedere la pagina 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 109 110

Commenti su questo manuale

Nessun commento