Agilent B1500 User’s Guide, Edition 7 5- 5
Classic Test Definition
I/V Sweep
Channel Setup
This setup screen is used to define the channel setups and the variables for the
channel output data, measurement data, time stamp data, and index data.
Channel Definition This area defines the channel setups and the v
ariables for th
e channel output data
and the measurement data.
Unit Module used for the source output or measurement
V Name Variable name of voltage output or measurement data
I Name Variable name of current output or measurement data
Mode Output mode. For SMU, V (voltage), I (current), VPULSE
(v
oltage pulse), IPULS
E (current pulse), or COMMON
(ground). For SPGU, VPULSE (voltage pulse) or ALWG
(arbitrary linear waveform voltage).
Function Output function, VAR1 (primary sweep), VA
R2 (secondary
sweep), VAR1’ (synchronous sweep), or CONST (constant)
Add SMU Adds a row for the SMU channel setup.
Add SPGU Adds a row for the SPGU channel setup.
Delete Deletes the channel setup. To select the channel setup, use the
l
eft radi
o button.
Up Moves the selected channel setup upward.
Down Moves the selected channel setup downward.
If multiple output channels are defined, the channels start output sequentially. The
o
rder is top to
bottom of the channels defined in this screen. The output stops in the
opposite order.
Miscellaneous
Va
riabl
es
This area defines the variables for the time stamp data and the index data.
Time Stamp Name Variable name of the time stamp. Time stamp data gives the
time when step measurem
ent is started.
Index Name Variable name of the data index. Integer above 0 is returned.