9-28 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
Error Codes
105804 Bias hold function execution error. Output voltage or voltage
compliance must be ± 40 V or less to control the output switch or the
series r
esistor.
105811 Invalid value A was specified for S
emiconductor Relays (B) of
Advanced Setup. Valid option(s): C
105821 Invalid value A was specified for T
ype of SPGU Pulse Setup. Valid
option(s): B, VName: C
105822 Invalid value A was specified f
or SW Sync of Pulse Switch Setup. Valid
option(s): B, VName: C
105823 Invalid value A was specified for No
rmal of Pulse Switch Setup. Valid
option(s): B, VName: C
105824 Width value of Pulse Switch Setup must be between A s and B s.
VName: C
1058
25 Width for pulse switch must be between A s and B s. VName: C
105826 S
um of Delay and Width of Pulse Switch Setup must be within Period
value. VName: A
1058
30 Load Z value must be between A ohm and B ohm. VName:
C
105831 Period value of SPGU Pulse Setup must be between A s and B s.
105832 PUL
SE COUNT value of SPGU Pulse Setup must be between A and B.
105833 DURAT
ION value of SPGU Pulse Setup must be between A s and B s.
1058
34 Base value of SPGU Pulse Setup must be between A V and B V.
VName: C
105835 Peak v
alue of SPGU Pulse Setup must be between A V and B V.
VName: C
105836 Delay
value of SPGU Pulse Setup must be between A s and B s.
VName: C
1058
37 Width value of SPGU Pulse Setup must be between A s and B s.
VName: C
1058
38 Leading value of SPGU Pulse Setup must be between A s and B s.
VName: C
1058
39 Trailing value of SPGU Pulse Setup must be between A s and B s.
VName: C
1058
40 Leading × 1.25 must be within Width value. VName: A