User’s Guide 27
E8362/3/4C Option H85
Example: Making High Power Measurements with Option H85
Final Setup
1. Verify that all of the power and attenuator settings are correct.
2. In the Response drop-down menu select Measure > S21.
Figure 14 Response Menu
3. To Perform a response calibration following steps a, b, c and d:
a. Connect the test port cables of the analyzer to form a thru configuration.
b. In the Response drop-down menu select Cal Wizard (see Figure 14). In the
application window, select UNGUIDED Calibration (Response, 1-port 2-port): Use
Mechanical Standards > Next.
Figure 15 Unguided Calibration