Communicating over the bus
Since GPIB can address multiple devices through the same interface card,
the device address passed with the program message must include not only
the correct interface select code, but also the correct instrument address.
Interface Select Code (Selects Interface)
Each interface card has a unique interface select code. This code is used by
the controller to direct commands and communications to the proper
interface. The default is typically “7” for GPIB controllers.
Instrument Address (Selects Instrument)
Each instrument on an GPIB must have a unique instrument address
between decimal 0 and 30. The device address passed with the program
message must include not only the correct instrument address, but also the
correct interface select code.
DEVICE ADDRESS = (Interface Select Code * 100) + (Instrument Address)
For example, if the instrument address for the oscilloscope is 4 and the
interface select code is 7, when the program message is passed, the routine
performs its function on the instrument at device address 704.
For the oscilloscope, the instrument address is typically set to
Oscilloscope Device Address
The examples in this manual and in the online Agilent 54600-Series
Oscilloscopes Programmer’s Reference assume the oscilloscope is at device
address 707.
See the documentation for your GPIB interface card for more information on
select codes and addresses.
Programming over GPIB
Communicating over the bus
3–5