Agilent Technologies HP 54601B Specifiche Pagina 91

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 100
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 90
Command Query Options and Query Returns
:SEQuence:SETup
{MASK | STEP}, <number>,
<header> <setup_string>
:SEQuence:SETup?
{MASK | STEP}, <number>
MASK ::= an individual mask sent to the setup string.
STEP ::= an individual step sent to the setup string.
<number> ::= the mask number or step number sent to the setup string.
<header> ::= the type of setup to be sent or returned:
For individual masks, ::= #800001000.
For individual steps, ::= #800000244.
For whole sequences, ::= #800064122.
<setup_string> ::= the setup string to be sent:
For individual masks, := 1000-byte string.
For individual steps, ::= 244-byte string.
For whole sequences, ::= 64122-byte string.
:SEQuence:STEP
<number>
:SEQuence:STEP? <number> ::= an integer from 1 to 100 in NR1 format
n/a :SEQuence:TEST? <result> ::= an integer from 0 to 3 in NR1 format:
0 = pass
1 = fail minimum limit line
2 = fail maximum limit line
3 = fail both minimum and maximum limit lines
*SRE
<mask>
*SRE? <mask> ::= sum of all bits that are set, 0,...,255; an integer in NR1
format. <mask> ::= following values:
Bit Weight Enables
7128Not Used
6 64 RQS - Request Service
5 32 ESB - Event Status Bit
4 16 MAV - Message Available
3 8 Not used
2 4 Not used
1 2 Not used
0 1 Not used
n/a :STATus?
<display>
{ON | OFF}
<display> ::=
{CHANnel <n> | PMEMory{1 | 2}} for the 54600/01/02/03/15/16
{CHANnel <n> | PMEMory{1 | 2} | EXTernal} for the 54610
<n> ::=
1 or 2; an integer in NR1 format for the 54600/03/10/15/16
1, 2, 3, or 4; an integer in NR1 format for the 54601/02
n/a *STB? <value> ::= 0,...,255; an integer in NR1 format, as shown in the following:
Bit Weight Name Condition
7 128 ---- NOT USED
6 64 RQS/MS 0 = instrument has no reason for service
1= instrument is requesting service
5 32 ESB 0 = no event status conditions occurred
1 = enabled event status condition occurred
4 16 MAV 0 = no output messages are ready
1 = an output message is ready
3 8 ---- 0 = not used
2 4 ---- 0 = not used
1 2 ---- 0 = not used
0 1 ---- 0 = not used
Programmers Quick Reference
Commands and Queries
8–14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Vedere la pagina 90
1 2 ... 86 87 88 89 90 91 92 93 94 95 96 ... 99 100