:SEQuence:TEST?
Query :SEQuence:TEST?
Return Format <Result><NL>
The :SEQUENCE:TEST query performs a mask template test and returns the
result of the testing.
<Result> An integer, 0 to 3 (NR1 format).
0= Pass.
1= Fail minimum limit line.
2= Fail maximum limit line.
3= Fail both minimum and maximum limit line.
Example 10 OUTPUT 707; ":SEQ:TEST?"
20 ENTER 707; Result
30 PRINT Result
40 END
When the fail condition is set to be inside, only 0 or 3 is returned.