7
Active Differential Probes
20 MHz to 1.5 GHz bandwidth
Switchable attenuation
Measure up to 600 V CAT III and
1000 V CAT II
N2772A 20 MHz differential
probe
Use the Agilent N2772A differential
probe with any of the 5000, 6000,
or 7000 Series oscilloscopes to
safely measure floating circuits with
the oscilloscope grounded. With
20 MHz bandwidth and switchable
attenuation of 20:1 and 200:1, it
provides the versatility for a broad
range of applications including
high-voltage circuits, motor controls,
power supply design, and electronic
high-power converters.
Each probe comes with 2 sharp probe
tips for use on small components and
in tight places, 2 retractable probe
hooks for connecting to smaller wires
and through-hole components, and
2-alligator clips for use with larger
cables.
This probe requires a 9 V battery or
Agilent N2773A power supply.
Specifi cations for N2772A differential
probe
Bandwidth 20 MHz
Risetime 17.5 ns
Attenuation ratio 20:1 and 200:1
selectable via switch
on probe
High CMRR 80 dB @ 60 Hz,
50 dB @ 1 MHz
Max input
voltage to ground
600 V CAT III
1000 V CAT II
Max differential
input voltage
1000 VDC or
1000 Vrms or
1200 VDC + peak AC
Measure up to 600 V CAT III
Max number
of probes
supported by
5000/6000/7000
Series
4
1141A 200 MHz differential
probe
The 1141A is a 1x FET differential
probe with 200-MHz bandwidth and
a 3000:1 CMRR. The probe has a
high-input resistance and low input
capacitance of 7-pF to minimize
circuit loading. The 1141A must be
used with 1142A probe control and
power module, which controls input
coupling modes dc, dc with variable
offset, and dc reject. Two attenuators,
10x and 100x are provided to expand
the linear differential input range to
±30V. This probe works with any 50-Ω
input oscilloscope including 5000 and
6000 Series.
Specifi cations for 1141A differential
probe
Bandwidth 200 MHz
Risetime
(calculated)
1.75 ns
Attenuation ratio 10:1 and 100:1 with
attenuater
High CMRR 3000:1 at 1-MHz
10:1 at 100-MHz
Input impedance Between inputs:
1 MΩ, 7 pF
Max input
voltage
200 Vdc + peak ac
(probe alone)
500 Vdc + peak ac
(with attenuator)
Max number
of probes
supported by
5000/6000/7000
Series
4
Ordering information for Agilent
differential probes and power supply
N2772A 20 MHz differential probe,
supplied with a retractable
hook, sharp probe tips and
alligator clips
N2773A Power supply for N2772A
1141A 200 MHz differential probe
1142A Probe control and power
module for 1141A
700
600
500
400
300
200
100
0
0.01 0.02 0.05 0.1 0.2 0.5 1
MHz
2 5 10 20 50 100
Vrms
Derating of each input for the N2772A
N2772 20 MHz differential probe
1141A 200 MHz differential probe
Vedere la pagina 6