Infiniium 9000 Series performance characteristics
Vertical: scope channels 9064A 9104A 9254A 9404A
Analog bandwidth (–3 dB) 50 Ω
1
600 MHz 1.0 GHz 2.5 GHz 4 GHz
1MΩ 500 MHz 500 MHz 500 MHz 500 MHz
Typical Rise Time / Fall 540 ps 253 ps 142 ps 85 ps
Time 10% to 90% at 50 Ω
Typical Rise Time / Fall 360 ps 174 ps 98 ps 59 ps
Time 20% to 80% at 50 Ω
Input channels DSO9000 – 4 analog
MSO9000 – 4 analog + 16 digital
Input impedance
1
50 Ω ± 2.5%, 1 MΩ ± 1% (11pF typical)
Input sensitivity
3
1 MΩ: 1 mV/div to 5 V/div
50 Ω: 1 mV/div to 1 V/div
Input coupling 1 MΩ: AC (3.5 Hz), DC
50 Ω:DC
Bandwidth limit 20 MHz on 1 MΩ input ; 500 MHz up to full scope bandwidth in increments of 500 MHz
Vertical resolution
2,3
8 bits, ≥12 bits with averaging
Channel-to-channel isolation DC to 50 MHz: 50 dB
>50 MHz to 2.5 GHz: 40 dB
>2.5 GHz to 4 GHz: 25 dB
DC gain accuracy
1,2,3
± 2% of full scale at full resolution on channel scale ± 5 ˚C from cal temp
Maximum input voltage
1
1 MΩ: 150V RMS or DC, CAT I
± 250 V (DC + AC) in AC coupling
50 Ω: 5 Vrms
Offset range Vertical sensitivity Available offset
1 MΩ 1 mV to <10 mV/div ± 2 V
10 mV to <20 mV/div ± 5 V
20 mV to <100 mV/div ± 10 V
100 mV to <1 V/div ± 20 V
1 V to 5 V/div ± 100 V
50 ±12 div or ±4V, whichever is smallest
1 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±5 °C from firmware calibration temperature. Input
impedance is valid when V/div scaling is adjusted to show all waveform vertical values within scope display.
2 Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.
3 50Ω input: Full scale is defined as 8 vertical divisions. Magnification is used below 10mV/div, full-scale is defined as 80 mV. The major scale settings are 5mV, 10mV, 20mV, 50
mV, 100mV, 200 mV, 500 mV, 1V.
1MΩ input: Full scale is defined as 8 vertical divisions. Magnification is used below 5mV/div, full-scale is defined as 40 mV. The major scale settings are 5mV, 10mV, 20mV, 50