Infiniium 9000 Series performance characteristics
Trigger: digital channels MSO Models
Threshold range (user defined) ±8.0 V in 100-mV increments
Threshold accuracy ±(100 mV + 3% of threshold setting)
Measurements and math
Waveform measurements (can be made on either min or zoom window with up to 10 simultaneous measurements
with statistics)
Voltage (scope channels) Peak-to-peak, minimum, maximum, average, RMS, amplitude, base, top, overshoot, V overshoot,
preshoot, V
preshoot, upper, middle, lower, crossing point voltage , pulse top, pulse base, pulse
amplitude
Time (digital channels) Period, frequency, positive width, negative width, duty cycle, delta time
Time (scope channels) Rise time, fall time, period, frequency, positive width, negative width, duty cycle, Tmin, Tmax, Tvolt,
channel-to-channel delta time, channel-to-channel phase , count pulses , burst width, burst period,
burst interval, setup time, hold time
Mixed (scope channels only) Area, slew rate
Frequency domain FFT frequency, FFT magnitude, FFT delta frequency, FFT delta magnitude
Level qualification Any channels that are not involved in a measurement can be used to level-qualify all timing
measurements
Eye-diagram measurements Eye height, eye width, eye jitter, crossing percentage, Q factor, and duty-cycle distortion
Measurement modes
Statistics Displays the mean, standard deviation, minimum, maximum range, and number of measurement values
for the displayed automatic measurements
Histograms (scope channels)
Source Waveform or measurement (histogram on measurement requires EZJIT, EZJIT+, or EZJIT Complete
option)
Orientation Vertical (for timing and jitter measurements) or horizontal (noise and amplitude change) modes,
regions are defined using waveform markers
Measurements Mean, standard deviation, mean ± 1, 2, and 3 sigma, median, mode, peak-to-peak, min, max, total
hits, peak (area of most hits), X scale hits, and X offset hits
Marker modes Manual markers, track waveform data, track measurements
Waveform math
Number of functions 16
Operators Absolute value, add, AM demodulation, average, horizontal gating, Butterworth
2
, common mode, dif
ferentiate, divide, FFT magnitude, FFT phase, FIR
11
, high pass filter, integrate, invert, LFE
2
, low pass
filter (4th-order Bessel Thompson filter),
magnify, max, min, multiply, RT Eye
2
, smoothing,
SqrtSumOfSquare
2
, square, square root, subtract, versus
Chartstate (MSO models), charttiming (MSO models)
Automatic measurements Measure menu access to all measurements, ten measurements can be displayed simultaneously
Multipurpose Front-panel button activates up to ten pre-selected or ten user-defined automatic measurements
Drag-and-drop Measurement toolbar with common measurement icons that can be dragged and dropped onto the
measurement toolbar displayed waveforms
FFT
Frequency range DC to 10 GHz (at 20 GSa/s) or 5 GHz (at 10 GSa/s)
Frequency resolution Resolution = sample rate/memory depth
Window modes Hanning, flattop, rectangular , Blackman Harris, Force
2 Requires MATLAB software.