Agilent Technologies N5230C Manuale Utente Pagina 20

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 41
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 19
16
Table 7.
Test Port Input (Continued)
Description Specification Typical
Options
140/240
Options
145/245
Options
146/246
Options
140/240
Options
145/245
Option
146/246
Trace Noise Phase
Ratioed measurement, nominal power at test port.
100 KHz IF bandwidth
300 KHz to 10 MHz -- -- -- 0.110 deg
rms
0.180 deg
rms
0.280 deg
rms
10 MHz to 10.5 GHz 0.05 deg
rms
0.07 deg
rms
0.13 deg
rms
0.025 deg
rms
0.035 deg
rms
0.090 deg
rms
10.5 GHz to 13.5 GHz 0.08 deg
rms
0.10 deg
rms
0.250 deg
rms
0.050 deg
rms
0.060 deg
rms
0.170 deg
rms
13.5 GHz to 20 GHz 0.08 deg
rms
0.10 deg
rms
0.250 deg
rms
0.050 deg
rms
0.060 deg
rms
0.170 deg
rms
600 KHz IF bandwidth
300 KHz to 10 MHz -- -- -- 0.110 deg
rms
0.180 deg
rms
0.300 deg
rms
10 MHz to 10.5 GHz -- -- -- 0.080 deg
rms
0.100 deg
rms
0.200 deg
rms
10.5 GHz to 13.5 GHz -- -- -- 0.120 deg
rms
0.160 deg
rms
0.430 deg
rms
13.5 GHz to 20 GHz -- -- -- 0.120 deg
rms
0.160 deg
rms
0.430 deg
rms
100 KHz IF bandwidth
Measured at Maximum Specified Power
300 KHz to 10 MHz -- -- -- 0.040 deg
rms
0.050 deg
rms
0.075 deg
rms
10 MHz to 2 GHz -- -- -- 0.007 deg
rms
0.012 deg
rms
0.013 deg
rms
2 GHz to 10.5 GHz -- -- -- 0.012 deg
rms
0.015 deg
rms
0.030 deg
rms
10.5 GHz to 13.5
GHz
-- -- -- 0.040 deg
rms
0.060 deg
rms
0.150 deg
rms
13.5 GHz to 20 GHz -- -- -- 0.040 deg
rms
0.060 deg
rms
0.150 deg
rms
Stability Magnitude
Stability is defined as a ratio measurement made at the test port.
300 KHz to 10 MHz -- -- -- +/-0.015 dB/°C
10 MHz to 2 GHz -- -- -- +/-0.010 dB/°C
2 GHz to 4 GHz -- -- -- +/-0.015 dB/°C
4 GHz to 13.5 GHz -- -- -- +/-0.020 dB/°C
13.5 GHz to 16 GHz -- -- -- +/-0.020 dB/°C
16 GHz to 19 GHz -- -- -- +/-0.025 dB/°C
19 GHz to 20 GHz -- -- -- +/-0.030 dB/°C