6
Digital Multimeter Specifi cations
[1]
± (% reading + % range)
Funtion Range Frequency, Test Current or Burden Voltage 1 year Tcal ± 5º C
DC Voltage 600.0 mV
6.000 V
60.00 V
600.0 V
0.3 + 0.08
0.3 + 0.08
0.3 + 0.08
0.3 + 0.08
AC Voltage 600.0 mV – 600.0 V 50 Hz – 1 kHz
1 kHz – 30 kHz
1.0 + 0.2
3.0 + 0.2
AC + DC Voltage 6.000 V – 600.0 V 50 Hz – 1 kHz
1 kHz – 30 kHz
1.0 + 0.2
3.0 + 0.2
Resistance 600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
60.00 MΩ
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
1.0 + 0.2
Capacitance 60.00 nF
600.0 nF
6.000 μF
60.00 μF
300.0 μF
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
Diode 1.000 V 0.5 mA 2.0 + 0.08
[1] For temperatures between 0 ºC to 18 ºC and 28 ºC to 50 ºC, add 0.1% of reading + 0.02% of range for every degree Celsius.
Cursors
Manually place readout of horizontal (X, DX) and vertical (Y, DY).
Waveform math CH1 + CH2, CH1 – CH2, CH2 – CH1
FFT
[1]
Window Rectangular, Hanning, Hamming, Blackman-Harris
Amplitude display Selectable in amplitude displays of 1 dB, 2 dB, 5 dB, 10 dB
Measurement System
Autoscale Finds and displays all active scope channels, sets edge trigger mode on
highest numbered channel, sets vertical sensitivity on scope channel.
Requires voltage > 20 mVp-p, 0.5% duty cycle and frequency > 100 Hz.
Automatic measurement Measurements continuously updated.
Voltage Peak-to-peak, maximum, minimum, amplitude, top, base, +overshoot,
–overshoot, preshoot, RMS, mean and one cycle mean.
Time Frequency, period, +width, –width, and +duty cycle and –duty cycle on
any channel.
Rise time, fall time, delay and phase shift.