1100 LC Modules Control Manual - 72 - 03-914947-30:Rev.6
<0.005 min (RiseTime<0.12 sec) (DataRate=13.74 Hz)
>0.005 min (RiseTime>0.12 sec) (DataRate=13.74 Hz)
>0.010 min (RiseTime>0.25 sec) (DataRate=13.74 Hz)
>0.025 min (RiseTime>0.5 sec) (DataRate=13.74 Hz)
>0.05 min (RiseTime>1 sec) (DataRate=6.87 Hz)
>0.1 min (RiseTime>2 sec) (DataRate=3.43 Hz)
>0.2 min (RiseTime>4 sec) (DataRate=1.72 Hz)
>0.4 min (RiseTime>8 sec) (DataRate=0.86 Hz)
Analogs Control
In the Offset field, enter the offset value to adjust the baseline.
In the Attenuation scrolling list, enter the attenuation value. The
available values are:
0.97 mAU full range, 1.95 mAU, 3.90 mAU, 7.81 mAU, 15.62
mAU, 31.25 mAU, 62.50 mAU, 125 mAU, 250 mAU, 500 mAU,
1000 mAU, 2000 mAU and 4000 mAU.