1100 LC Modules Control Manual - 84 - 03-914947-30:Rev.6
Excitation Parameter Group
Check Wavelength A to Wavelength D to activate the
corresponding wavelengths in multi--wavelength mode. Enter
their values in the corresponding fields.
Check Zero Order to work at zero order.
Emission Parameter Group
Check Wavelength A to Wavelength D to activate the
corresponding wavelengths in multi--wavelength mode. Enter
their values in the corresponding fields.
Check Zero Order to work at zero order.
Acquire Spectra Parameter Group
Using the scrolling list, select the spectral Acquisition Mode.
None: no spectra are stored.
Apex: only the top of peak spectrum is stored.
All in peak: all spectra between the beginning and
the end of a peak are stored.
All: all spectra are stored during the run.
In the fields defining the Range, enter the minimum and
maximum wavelength of the spectrum. Enter then the
wavelength Step in the corresponding field. A small wavelength
step will enhance the spectrum resolution.
Enter the value of the Threshold in the corresponding field.