Agilent Technologies 1100 Series Manuale delle Istruzioni Pagina 26

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 272
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 25
26
Installing the Pump
Flow Connections of the Quaternary Pump
5 Connect the solvent tubes to the outlet connectors (typically right connection
of the channel) of the vacuum degasser.
6 Connect the syringe adapter from the degasser accessory kit to the solvent
tube of channel A.
7 Using a piece of sanding paper connect the waste tubing to the purge valve
and place it into your waste system.
8 If the quaternary pump is not part of a Agilent1100 System stack or placed on
the bottom of a stack, connect the corrugated waste tube to the waste outlet
of the pump leak handling system.
9 Connect the pump outlet capillary (quaternary pump to injection device) to
the outlet of the purge valve.
10 Prime your system before first use (see Priming and Purging the System on
page 28).
Vedere la pagina 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 271 272

Commenti su questo manuale

Nessun commento