Agilent Technologies 1100 Series Manuale delle Istruzioni Pagina 50

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 272
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 49
50
Troubleshooting and Test Functions
Remote Timeout
Remote Timeout
A not-ready condition is still present on the remote input .
When an analysis is started, the system expects all not-ready conditions (e.g.
a not-ready condition during detector balance) to switch to run conditions
within one minute of starting the analysis. If a not-ready condition is still
present on the remote line after one minute the error message is generated.
Probable Causes Not-ready condition in one of the instruments connected to the remote
line.
Defective remote cable.
Defective components in the instrument showing the not-ready condition.
Suggested Actions Ensure the instrument showing the not-ready condition is installed
correctly, and is set up correctly for analysis.
Exchange the remote cable.
Check the instrument for defects (refer to the instruments reference
documentation).
Vedere la pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 271 272

Commenti su questo manuale

Nessun commento