216 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
24 Calibrating the Infiniium Oscilloscope and Probe
3 Referring to Figure 81 below, perform the following steps:
a Click on the Utilities>Calibration menu to open the Calibration
window.
4 Referring to Figure 82 below, perform the following steps to start the
calibration:
a Uncheck the Cal Memory Protect checkbox.
b Click the Start button to begin the calibration.
Figure 81 Accessing the Calibration Menu.
Vedere la pagina 215