228 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
24 Calibrating the Infiniium Oscilloscope and Probe
2 Referring to Figure 93 below, perform the following steps:
a Click the Add Head... button, and then select N5380A:DF SMA from
the list of Head Type. Select OK to close the dialog box.
b Click on the Calibrate Probe button to open the Probe Calibration
window.
3 Referring to Figure 94 below, perform the following steps:
a Select the Calibrated Atten/Offset Radio Button
b Click the Start Atten/Offset Calibration Button to open the
Calibration window.
Figure 93 Probe Setup Window.
Vedere la pagina 227