DisplayPort Test Implementation Group 18 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
c. Alternate between adjusting PJ 1 on the J-BERT N4903B user interface and pressing the
Single Acquisition Button on the Oscilloscope until the PJ
(d-d)
is calibrated to the following
values.
For RBR
For HBR
For HBR2
NA
NA
Adjust PJ 1 to 100mUI as shown
in the below example
Repeat measurement at least 3
times and average the results.
Figure 12: J-BERT N4903B HBR2 SJ
FIXED
Measurement
Note: In this example, the measured PJ value is 101.2 mUI.
9.3.2 SJ
SWEEP
1. Calibrate SJ
SWEEP
at each frequency with JMD using constant clock at TP3_EQ. Note: To account for
small inaccuracies in measurement of ISI, RJ and SJ
FIXED
compared to the spec, the SJ
SWEEP
should be
adjusted as follows to ensure correct TJ at TP3_EQ:
For HBR: SJ
SWEEP
= SJ
SWEEP
(Table 4-3)*(1 + (323mUI - measured_mUI(ISI + RJ))/168mUI).
For HBR2: SJ
SWEEP
= SJ
SWEEP
(Table 4-4)*(1 + (520mUI - measured_mUI(ISI + RJ + SJ
FIXED
))/100mUI).
Notes:
RJ in this formula is Pk-Pk = 12.0 x RJ(rms).
In the above equations, the denominator of the measurement error correction (168mUI
for HBR and 100mUI for HBR2) is equivalent to the high frequency SJ
SWEEP
value in