Agilent Technologies DisplayPort Manuale Utente Pagina 22

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 56
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 21
Agilent Technologies, Inc.
DisplayPort Test Implementation Group 22 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 18: HBR2 crosstalk Setup for Trigger/Reference Clock Output
e. In the J-BERT N4903B go to PG Setup > Clock/Jitter Output menu.
i. Set the Trigger Output to the following based on the data rate being testing:
Figure 19: Controlling Crosstalk Amplitude on the J-BERT N4903B
For RBR
For HBR
For HBR2
Set crosstalk amplitude to
414mV
diff
.
Set cross talk amplitude to
1.35V
diff
.
Set the Clock/Trigger output
amplitude to 0V during this
step. Crosstalk amplitude will
be applied in 9.4.2.
3. If the Sink DUT supports FAUX, ensure that the DUT is transmitting FAUX traffic to emulate
crosstalk during calibration.
Vedere la pagina 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 55 56

Commenti su questo manuale

Nessun commento