DisplayPort Test Implementation Group 24 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 21: J-BERT N4903B Differential Amplitude Adjustment
b. In the oscilloscope, using the Analysis > Serial Data > Clock Recovery menu, change the
Clock Recovery from Constant Frequency to 2
nd
Order PLL.
Figure 22: DSA 2
nd
Order PLL Clock Recovery
c. On the Oscilloscope, Select Measure > Eye > Eye Height Measurement.
d. Alternate between adjusting the Vampl on the J-BERT N4903B user interface and pressing
the Single Acquisition Button on the Oscilloscope until the Eye Height is calibrated to the
following values.
Adjust Eye Height to 150mV
diff
+/-10%
Repeat measurement at least 5
times and average the results.
Adjust Eye Height to 120mV
diffI
+/-10% as shown in the below
example
Repeat measurement at least 5
times and average the results.