Setting Up a Repetitive Burst Signal Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 53
The instrument panel now looks as follows:
Channel Settings for the PRBS Signal To set the channel parameters:
1 Switch to the Channel page.
2 In the Timing panel:
Set the signal mode to NRZ.
Set the Delay to 50 ps.
3 In the Levels panel, set:
The level format to Custom.
Amplitude to 2.0 V.
Offset to 0.0 V.
4 Enable the Channel 1 normal Output:
By pressing the respective softkey in the user interface:
– OR –
By pressing the normal Out softkey for channel 1 next to the
Output port.
Burst Repetition
Instrument
Mode
Specifies the
number of zeros
Vedere la pagina 52