Setting Up a Repetitive Burst Signal Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 55
Reconnect the DUT Enable the outputs:
Generated Signal Use the pulse generator’s TRIGGER OUT to trigger a scope.
To set up the trigger output:
1 On the Aux page, in the Trigger Output panel:
Select the trigger mode Pulse.
One trigger pulse occurs at the start of the repetitive data
pattern.
Set the Ampl to 1 V.
Set the Offset to 0 V.
Set the Divider to 256.
2 Enable the trigger output.
The Aux page looks now as follows: