Front panel connectors
Connector type type-N female
Impedance 50 ohms (standard)
75 ohms (Option 1EC)
Probe power +15V 200 mA
-12.6V 250 mA
Rear panel connectors
External reference 10 MHz, > –5 dBm,
50 ohm BNC
Auxiliary input
The auxiliary input measures the DC level at each
sweep point. If the slew rate on this input exceeds
700 mV/msec, increased measurement errors will
result.
Calibrated range ±10V
Accuracy ±(3 % of reading +20 mV)
Damage level >15 Vdc
External trigger
Triggers on a negative TTL transition or contact
closure to ground.
Limit test output
Provides an open collector TTL high signal. The
output is pulled low when the limit test fails.
User TTL input/output
Provides a bi-directional open collector TTL signal
that can be accessed by IBASIC.
Video output
Provides an RS-343A compatible multisync video
signal. Pixel rate is 33.3 MHz, vertical rate is 60 Hz,
and horizontal rate is 24.1 kHz. Output is not com-
patible with EGA or VGA monitors.
GPIB
Allows communications with compatible devices
including external controllers, printers, plotters,
disk drives, and power meters.
X and Y external detector inputs
Provides for two external detector inputs. See
Agilent 86200B and 86201B Data Sheet, literature
number 5962-9931E.
Parallel port
This 25-pin female connector is used with parallel
(or Centronics interface) peripherals such as print-
ers and plotters. It can also be used as a general-
purpose I/O port, with control provided by IBASIC.
RS-232C
This 9-pin male connector is used with serial
peripherals such as printers and plotters.
DIN keyboard
This connector is used for adding an IBM PC-AT
compatible keyboard for titles, remote front-panel
operation, and for IBASIC programming (Option
1C2).
Liner
47 to 60 Hz
115V nominal (90V to 132V) or 230V nominal
(198V to 264V) 230 VA max. A three-wire ground is
required.
Environmental Characteristics
General conditions
RFI and EMI susceptibility defined by CISPR
Publication 11.
ESD (electrostatic discharge) should be minimized
by the use of static-safe work procedures and an
antistatic bench mat (such as a 92175T).
The sealed flexible rubber keypad protects key con-
tacts from dust, but the environment should be as
dust-free as possible for optimal reliability.
Operating environment
Temperature to 55° C
Humidity 5% to 95% at 40° C
(noncondensing)
Altitude 0 to 4,500 meters (15,000 feet)
Storage conditions
Temperature –40° C to +70° C
Humidity 0 to 90% relative at +65° C
(noncondensing)
Altitude 0 to 15,240 meters (50,000 feet)
Cabinet dimensions
The following dimensions exclude front and rear
panel protrusion:
179 mm H x 425 mm W x 514 mm D
(7.0 in x 16.75 in x 20.25 in)
Weight
Net 20.5 kg
Shipping 30 kg
10