5
Source Signal Purity
Nonharmonic spurious Agilent 8712B Agilent 8714B
≥50 kHz from carrier <–20 dBc, <1 MHz <–30 dBc
<–30 dBc, ≥1 MHz
<50 kHz from carrier <–25 dB <–25 dBc
Phase noise –70 dBc/Hz –67 dBc/Hz
(at 10 kHz offset)
Residual AM <–50 dBc <–50 dBc
(in 100 kHz bandwidth)
Residual FM <1.5 kHz <1.5 kHz
30 Hz to 15 kHz peak peak
Display Characteristics
Amplitude
Display resolution 0.01 dB/division
Marker reference level range: ±500 dB
resolution: 0.01 dB
Phase
Range ±180°
Display resolution 0.1°/division
Marker resolution 0.01°
Reference level range ±360°
resolution 0.01°
Polar scale range 1m to 20/division
Typical measurement uncertainty for 8714B at 1.3 GHz
Transmission magnitude uncertainty
Transmission phase uncertainty
Reflection magnitude uncertainty
Reflection phase uncertainty
These graphs show the measurement uncertainty for the 8714B. The assumptions made to generate these curves were: for transmission
uncertainty, S
1
1
= S
22
= 0.0; and for the reflection uncertainty, S21 =S12 = 0.0. Reflection tracking = 0.3 dB, transmission tracking =
0.2704 dB (computed from match terms), and trace noise = 0.250 dB. Power = 0 dBm for reflection measurements, and –20 dBm for
transmission measurements.
Supplemental Data