Agilent Technologies A.18.00 Manuale Utente Pagina 211

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 581
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 210
211
IEEE 488.2 Common Commands
*RST (Reset) The *RST command resets the Test Set. When the *RST command is received the
majority of fields in the Test Set are “restored” to a default value, some fields are
“maintained” at their current state and some are “initialized” to a known state.
Other operational characteristics are also affected by the *RST command as
follows:
All pending operations are aborted.
The Test Set’s display screen is in the UNLOCKED state.
Measurement triggering is set to TRIG:MODE:SETT FULL;RETR REP.
Any previously received Operation Complete command (*OPC) is cleared.
Any previously received Operation Complete query command (*OPC?) is cleared.
The power-up self-test diagnostics are not performed.
The contents of the SAVE/RECALL registers are not affected.
Calibration data is not affected.
The GPIB interface is not reset (any pending Service Request is not cleared).
All Enable registers are unaffected: Service Request, Standard Event, Communicate,
Hardware #1, Hardware #2, Operation, Calibration, and Questionable Data/Signal.
All Negative Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operational, Calibration, and Questionable Data/Signal.
All Positive Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operational, Calibration, and Questionable Data/Signal.
The contents of the RAM memory are unaffected.
The contents of the Output Queue are unaffected.
The contents of the Error Queue are unaffected.
Vedere la pagina 210
1 2 ... 206 207 208 209 210 211 212 213 214 215 216 ... 580 581

Commenti su questo manuale

Nessun commento