Chapter 8, Programming the Call Processing Subsystem
Programming the CALL CONFIGURE Screen
Programming the CALL CONFIGURE Screen
Figure 57 The CALL CONFIGURE Screen
This screen is used to set some of the less commonly used Test Set configuration
parameters.
When the CALL CONFIGURE screen is displayed and the Call Processing
Subsystem is in the
Connect state, the host firmware constantly monitors the
mobile station’s transmitted carrier power. If the power falls below 0.0005Watts
the error message
RF Power Loss indicates loss of Voice Channel
will be displayed and the Test Set will terminate the call and return to the
Active
state.