Chapter 8, Programming the Call Processing Subsystem
Programming the CALL CONTROL Screen
Programming the CALL CONTROL Screen
Figure 34 The CALL CONTROL Screen
The CALL CONTROL screen is the primary Call Processing Subsystem screen. It
contains the most often used Test Set configuration fields and the command fields
used to initiate call processing functions.
[] Access
When lit, the Access annunciator indicates that the Test Set is signaling the
mobile station with command information on the forward voice channel. This is a
transitory state.
The
Access annunciator is not programmable.
The state of the
Access annunciator is reflected in the Call Processing Status
Register Group Condition Register bit 4. See “Status Reporting” in the User’s
Guide for further information.
The Test Set’s speaker is turned off when in the Access state. This is done to
eliminate any possible audio feedback which may occur if the mobile station’s
microphone is open.