Agilent Technologies InfiniiVision 3000 3024A MSO-X Specifiche Pagina 12

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 37
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 11
12 | Keysight | InniiVision 3000T X-Series Oscilloscopes - Data Sheet
Solve: Integrated hardware-based serial decoding and triggering
(option) makes easy work of low speed serial buses
Keysight InniiVision oscilloscopes, including the new 3000T X-Series, use
hardware-based serial protocol decoding. Some other vendors use software
post-processing techniques to decode serial packets⁄ frames, and therefore have slow
waveform and decode capture rates and could miss critical events and errors due to a
long dead-time. Faster decoding with hardware-based technology enhances the
probability of capturing infrequent serial communication errors.
After capturing serial bus communication, you can easily perform a search operation
based on specic criteria and then quickly navigate to bytes/frames of serial data that
satisfy that search criteria. The 3000T X-Series can decode two serial buses
simultaneously using hardware based decoding, and display the captured data in a time
interleaved “lister” display.
Serial protocol decoding can be used simultaneously with segmented memory and
Zone touch triggering. The 3000T X-Series has the most decode/trigger
options in this class of instrument including: I²C, SPI, RS232/422/485/UART, CAN,
CAN-FD, CAN-dbc, LIN, SENT, FlexRay, MIL-STD 1553, ARINC 429, and I²S.
Figure 15: I
2
C decode and
trigger
Figure 16: RS232 decode
and trigger
Figure 17: CAN-FD decode
and trigger
Figure 18: SPI 4wire
decode and trigger
Figure 19: Multi-bus time
aligned decode
SERIAL DECODE AND
TRIGGER OPTIONS
The 3000T X-Series supports a
range of different serial decode and
trigger options including:
I
2
C
SPI (2/3/4 wire)
RS232/422/485/UART
CAN
CAN-dbc
CAN-FD
LIN
SENT
FlexRay
MIL-STD 1553
ARINC 429
I
2
S
Solve: Segmented smart memory combined with protocol analysis
enables insights over long periods of time
Segmented memory works in conjunction with any of the optional serial protocol
decodes. For example, by setting the trigger condition to “SENT serial bus error,
segmented memory captures and stores only SENT pulse period error packets and
stitches together each segment for easy viewing of the decoded data in the lister. You
can quickly compare time tags to discover time intervals between errors.
Figure 20: Segmented memory being used in conjunction with serial decode resulting in maximum