17 | Keysight | InniiVision 3000T X-Series Oscilloscopes - Data Sheet
Integrated frequency measurements: Optional 8-digit counter and
totalizer
Traditional oscilloscope counter measurements offer only ve or six digits of resolution,
which may not be enough for the most critical frequency measurements are being made.
With the 3000T X-Series’ optional 8-digit counter (part of the DSOXT3DVMCTR option),
you can see your measurements with the precision you would normally expect only from
a standalone counter. Because the integrated counter measures frequencies up to a wide
bandwidth of 1.0 GHz, you can use it for many high-frequency applications as well. If you
are looking for the ultimate precision, you can connect your 3000T X-Series oscilloscope
to your most trusted 10-MHz reference source to share a common 10-MHz clock.
The totalizer feature of the DSOXT3DVMCTR counter option adds another valuable
capability to the oscilloscope. It can count the number of events (totalize), and it also can
monitor the number of trigger-condition-qualied events. The trigger-qualied events
totalizer does not require an actual trigger to occur. It only requires a
trigger-satisfying event to take place. In other words, the totalizer can monitor events
faster than the trigger rate of a scope, as fast as 25 million events per second (a
function of the oscilloscope’s holdoff time, which has the minimum of 40 ns). Figures 27
shows example of a totalizer counting the number of CAN-FD CRC delimiter bit error
packets that took place in a design.
See www.keysight.com/nd/DSOXDVMT3CTR for more information
Figure 27: Totalizer counts the number of events. In addition, it can count the number of
trigger-condition-qualied events as fast as 25 million events a second.