Chapter 5: Replacing Assemblies
To set the calibration factors after replacing the acquisition board
114
c Pull down the Utilities menu and select Calibration.
d Uncheck the Cal Memory Protect box to allow calibration.
e Click on Start to start the calibration procedure.
Follow the on-screen instructions as calibration proceeds.
Vedere la pagina 113