Chapter 3: Testing Performance
Offset Accuracy Test
26
b Pull down the Setup menu and select Acquisition. When the Acquisition Setup window
is displayed, enable averaging and set the # of averages to 256.
c Change the vertical sensitivity of channel 1 to 5 mV/div.
d Click the V avg measurement icon on the left side of the screen.
e When the Enter Measurement Info window is displayed, ensure that the V avg function
is set up as follows and then click OK:
Source = Channel 1
Measurement area = Entire Display
Vedere la pagina 25