Chapter 5: Replacing Assemblies
To remove and replace the cover, top plate, and bottom plate
94
T10 screws from its edges as shown in Figure 5-6. Then pull back on the top plate and
lift off as shown in Figure 5-7.
Figure 5-6
Figure 5-7
10
To remove the bottom plate, flip the oscilloscope upside down and remove the three
Torx T20 screws from the bottom panel as shown in Figure 5-8. Then remove the three
Torx T20 screws from the bottom edge of the rear panel as shown in Figure 5-9. Slide
the bottom panel up and out (Figure 5-10).