Agilent Technologies N5230C Manuale di Servizio Pagina 59

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 58
Service Guide N5230-90025 3-27
PNA Series Microwave Network Analyzers Tests and Adjustments
N5230C System Verification
Figure 3-12 shows an example of typical verification results with Print Graphs selected in the Printer Output
area of the System Verification dialog box. The printed graphical results show the following:
the name of the device measured
the serial number of the device
the parameters measured
Results of the measurements. Labeled as A in Figure 3-12.
Data measured at the factory from the verification kit. Labeled as B in Figure 3-12.
Upper and lower limit points as defined by the total system uncertainty system. Labeled as C in Figure
3-12.
Figure 3-12 Example of Printed Graphical Verification Results
Vedere la pagina 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 295 296

Commenti su questo manuale

Nessun commento