Agilent Technologies E1445A Guida Utente Pagina 304

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 539
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 303
4-68 Agilent E6432A Microwave Synthesizer User’s Guide
HPE6432_GetFreqModExtSensitivity
ViStatus HPE6432_GetFreqModExtSensitivity (ViSession
instrumentHandle, ViReal64 *HZperVolt);
Purpose
This function gets the current FM sensitivity setting.
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
HZperVolt
Variable Type ViReal64 (passed by reference)
This parameter returns the FM sensitivity specified by the Set FM
External Sensitivity function.
Allowable values: 10E6, 1E6, 100E3
These values correspond to 10 MHz/V, 1 MHz/V, or 100 KHz/V.
Return Value
This return value always returns VI_SUCCESS.
Related Topics
HPE6432_SetFreqModExtSensitivity
Vedere la pagina 303