Agilent Technologies E1445A Guida Utente Pagina 66

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 539
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 65
1-52 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 6. External Leveling
Test 6. External Leveling
Description
The E6432A microwave synthesizer is configured for external leveling
using a negative detector and a directional coupler. An E6432A
microwave synthesizer user calibration is then performed and the
synthesizer is checked for leveling.
Required Test Equipment
33330C detector
NOTE A different negative detector may be used depending upon the
frequency range of the external leveling loop configuration.
Krytar directional coupler with 16 dB coupling factor
EPM-441A (E4418A) single channel power meter
ECP-E26A (E4413A) 50 MHz to 26.5 GHz power sensor
Vedere la pagina 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 538 539

Commenti su questo manuale

Nessun commento