4-244 Agilent E6432A Microwave Synthesizer User’s Guide
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
startingPoint
Variable Type ViUInt32
This parameter specifies the first point to write to list point memory.
Valid range: 0 to (131,070 - Number_of_Points)
Factory Preset Value: 0
frequency
Variable Type ViReal64[]
This parameter specifies an array of frequency settings.
Valid Values: 10 MHz to 20 GHz with 1 Hz resolution.
alcPower
Variable Type ViReal64[]
This parameter specifies an array of ALC power level settings.
Valid range: –20 dBm to Maximum Leveled Output Power
attenuation
Variable Type ViInt16[]
This parameter specifies an array of attenuation settings.
Valid values: 0 to 70 dB in 10 dB steps
50,000 0.46 0.46 0.55 6.92
100,000 0.92 0.92 1.11 13.51
131,071 1.16 1.20 1.38 17.62
Number
of Points
PII-400/MXI-2 PII-400 P-150/MXI-2 P-90