4-92 Real-Time I/Q Baseband Generator User’s and Programming Guide
Programming Commands and Examples ESG Family Signal Generators
Programming Examples Options UN8 and 202
PHS
Use this example to configure the signal generator to the initial settings of an PHS
π/4DQPSK modulation sensitivity bit error test. The carrier frequency is 1.89515 GHz at
40 dBuV (displayed in dBm). You can change the PHS format default values once the setup
is complete.
CLEAR and RESET the controller, type the following commands and RUN the program.
10 !********************************************************************
20 !
30 ! PROGRAM NAME: PHS Rev. 2A796
40 !
50 ! PROGRAM DESCRIPTION: In this example, the instrument is configured to
60 ! the initial settings of a PHS Pi/4 DQPSK modulation sensitivity
70 ! Bit error test. The carrier is a frequency of 1.89515 GHz
80 ! at 40 dBuV (displayed in dBm). The PHS format is
90 ! initially set with default values which can be changed
100 ! once the setup is complete.
110 !
120 !
130 !
140 !
150 !********************************************************************
160 !
170 Sig_gen=719
180 LOCAL Sig_gen
190 CLEAR Sig_gen
200 CLEAR SCREEN
210 OUTPUT Sig_gen;"*RST"
220 OUTPUT Sig_gen;"*CLS"
230 ! *********************************************
240 OUTPUT Sig_gen;"FREQ 1.89515 GHz"
250 OUTPUT Sig_gen;"POW 40 dBuV"
260 OUTPUT Sig_gen;"RAD:PHS:BURS:STAT ON"
270 OUTPUT Sig_gen;"RAD:PHS:DLIN:SLOT1:TYPE TCH"
280 OUTPUT Sig_gen;"RAD:PHS:DLIN:SLOT1:STAT ON"
290 ! *********************************************
300 OUTPUT Sig_gen;"RAD:PHS:STAT ON"
310 ! *********************************************
320 OUTPUT Sig_gen;"OUTP ON"
330 ! *********************************************
340 LOCAL Sig_gen
350 PRINT
360 PRINT "Wait until the `Baseband Data Generation' is complete, then ..."
370 PRINT
380 WAIT 5
390 PRINT "... Press the `Amplitude' key to enable the UP/DWN arrows"
400 PRINT "and RPG knob in order to adjust the output power level."
410 PRINT
420 WAIT 5
430 PRINT "Press RUN to start again."
440 END
Vedere la pagina 335