Chapter 6 147
8662A/63A Commands
8662A/63A Compatible Commands
SS|BLST Set sequence N N not supported
ST
Store
Saves/recalls register to sequence 0.
YY*SAV
T1
0.5 ms per step
YY
SWEEP:DWELL 0.5ms
Beyond ESG range limit; is set to 1ms.
T2 1 ms per step Y Y SWEEP:DWELL 1ms
T3 2 ms per step Y Y SWEEP:DWELL 2ms
T4 10 ms per step Y Y SWEEP:DWELL 10ms
T5 100 ms per step Y Y SWEEP:DWELL 100ms
TR
Trigger
Performs command code setup with CT command.
YY
no equivalent SCPI command sequence
UP
Increment
Passes UP as a parameter of the active function command.
YY
See Table 6-2 on page 148
UV mV Y Y UV
W1 Sweep off Y Y
FREQ:MODE CW
LIST:TRIG:SOUR IMM
W2
Auto sweep mode on
Generates a sweep list based on stored parameters from
FA, FB, FR, FS, N1, N2, N3, N4, and N5
Default values:
FR = 100 MHz,
FS = 10 MHz,
N1, T2 FA = 1 MHz,
FB = 1279 MHz
YY
INIT:CONT ON
SWEEP:MODE AUTO
LIST:TRIG:SOUR IMM
LIST:DWELl:TYPE STEP
LIST:TYPE LIST
FREQ:MODE LIST
W3
Manual sweep mode on
Generates a sweep list based on stored parameters from
FA, FB, FR, FS, N1, N2, N3, N4, and N5
Default values:
FR = 100 MHz,
FS = 10 MHz,
N1, T2 FA = 1 MHz,
FB = 1279 MHz
YY
INIT:CONT ON
SWEEP:MODE MANual
LIST:TRIG:SOUR IMM
LIST:DWELl:TYPE STEP
LIST:TYPE LIST
FREQ:MODE LIST
Table 6-1 8662A/63A Commands & Equivalent SCPI Sequences (Continued)
Command Description 8662 8663 Equivalent SCPI Command Sequence