Chapter 2 11
E4428C/38C SCPI Commands
Digital Modulation Subsystem (:SOURce)
I/Q skew adjustments are preserved when the instrument state is saved. I/Q skew adjustment are also
preserved when instrument settings are changed. If the signal generator is calibrated, the skew adjustments
are added to the calibration value used for the given signal generator state. If the signal generator is
uncalibrated, the skew adjustments re applied directly.
Using I/Q skew while playing a user FIR file greater than 32 symbols will generate an error.
The variable <val> is expressed in units of picoseconds or nanoseconds.
*RST +0.00000000E+000
Range 5.0 to 5.0
Key Entry
I/Q Skew
Remarks This command is no longer supported and has been replaced by:
[:SOURce]:DM:IQADjustment:SKEW[:DELay] <val>