Chapter 13 197
SCPI Command Reference
IEEE Common Commands
13. SCPI Command
Reference
*ESE
Syntax *ESE <numeric>
*ESE?
Description Sets the value of the Standard Event Status Enable Register.
Parameters
If the specified parameter is out of the allowable setup range, the result of bitwise AND
with 255 (0xff) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;"*ESE 16"
20 OUTPUT 717;"*ESE?"
30 ENTER 717;A
Related commands *SRE on page 200
Equivalent key No equivalent key is available on the front panel.
*ESR?
Syntax *ESR?
Description Reads out the value of the Standard Event Status Register. Executing this command clears
the register value. (Query only)
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;"*ESR?"
20 ENTER 717;A
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Setup value of the register
Range 0 to 255
Preset value 0
Resolution 1
Vedere la pagina 196