66 Chapter 4
Performing a Calibration
Calibration
Saving and Loading Calibration Coefficients
You can save calibration coefficients to a file along with other instrument settings, and can
later load them from the file.
NOTE You cannot save, load, read, or write calibration coefficients independently of the other
instrument settings.
By default, the system does not save calibration coefficients when it saves instrument
settings. To save calibration coefficients, therefore, you must explicitly configure the
system to save calibration coefficients by issuing the following command:
:MMEM:STOR:STYP on page 363
To save calibration coefficients to a file, use the following command:
:MMEM:STOR on page 353
To load calibration coefficients from a file, use the following command:
:MMEM:LOAD on page 345
For more information on how to save and load calibration coefficients, refer to “Saving and
recalling instrument status.”