Chapter 6: Replacing Assemblies
To replace the module
1 If the module consists of one card, go to step 2.
If the module consists of more than one card, connect the cables together in a
master/expander configuration. Follow the procedure "To configure a multicard
module" in chapter 2.
2 Slide the cards above the slots for the module about halfway out of the
mainframe.
3 With the probe cables facing away from the instrument, slide the module
approximately halfway into the mainframe.
4 Slide the complete module into the mainframe, but not completely in.
Each card in the instrument is firmly seated and tightened one at a time in step 6.