Chapter 3: Testing Performance
To Test the Single-clock, Multiple-edge, State Acquisition
4 Set up the Trigger tab.
a In the Analyzer setup window, select the Trigger tab. Under the Trigger
tab, select the Settings tab at the bottom of the window.
b Select the Acquisition Depth field, then select “8K”.
c Select the Trigger Position field, then select Start.
Vedere la pagina 85