Chapter 3: Testing Performance
To Test the Threshold Accuracy
3 Using the Modify down arrow on the function generator, decrease offset
voltage in 1-mV increments until all activity indicators for the pod under
test show the channels are at a logic low. Record the function generator
voltage in the performance test record.