Agilent-technologies Multiple Channel Controller 85330A Manuale Utente Pagina 116

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 115
7-6 85330A Multiple Channel Controller
Remote Programming Command Reference
Standard Command Reference
ROUTe:CONTrol
Controls the TTL state of the rear panel AUX 1 connector.
Default Default is a TTL LOW.
Syntax OUT:CONT <data>;
Example ROUT:CONT 0;
Sets AUX 1 to TTL LOW
ROUT:CONT 1;
Sets AUX 1 to TTL HIGH
NOTE The state of AUX 1 must be set before entering run-time mode.
ROUTe:DELay
Adds a specific amount of delay between a switch address entry being
asserted and the issuance of a measurement trigger. Delay is entered in
micro-seconds.
Default The default is 160 micro-seconds.
Syntax ROUT:DEL <micro-seconds>;
Example ROUT:DEL 200;
Set the trigger delay (settling time) to 200 micro-seconds.
NOTE Most solid state switches are capable of changing states within a few
micro-seconds.
Vedere la pagina 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 165 166

Commenti su questo manuale

Nessun commento